AG Associates Heatpulse 4100 Utility Specification:
The following table is a summary of the utility specifications for the Heatpulse 8108 system:
UTILITY | Flow Rate | Pressure | ||||
Tube Cooling CDA or Utility N2 | 8 SCFM | 10 SCFM | 15 SCFM | 30 psi | 60 psi | 100 psi |
Valve Actuation CDA or Utility N2 | <1 SCFM | <1 SCFM | <1 SCFM | 70 psi | 80 psi | 100 psi |
Cooling Water Supply | 1.6GPM | 2.0 GPM | 2.5GPM | 30 psi | 40 psi | 100 psi |
Cooling Water Return | 1.6 GPM | 2.0 GPM | 2.5 GPM | 20 psi | 30 psi | 40 psi |
Gas Box Exhaust | 0 SCFM | 150 SCFM | >200 SCFM | .5″ H2O | .75″ H2O | ——– |
Cooling Exhaust | 8 SCFM | 10 SCFM | 15 SCFM | .5″ H2O | .75″ H2O | ——– |
Scavenger Hood Exhaust | 20 SCFM | 25 SCFM | 30 SCFM | .5″ H2O | .75″ H2O | ——– |
Containment Exhaust (Optional) | 0 SCFM | 0 SCFM | 2 SCFM | .5″ H2O | .75″ H2O | ——– |
Exhaust Manifold Outlet | 28 SCFM | 35 SCFM | 45 SCFM | .5″ H2O | .75″ H2O | ——– |
Process Gas Exhaust (Scrubber) | 10 SCFM | 10 SCFM | 10 SCFM | .5″ H2O | .75″ H2O | 2.5″ H2O |
Process Gas Supply | 10 SCFM | 10 SCFM | 10 SCFM | 30 psi | 40 psi | 60 psi |
Recirculator Water Supply | 3 GPM | 5 GPM | 7 GPM | 10 psi | 20 psi | 40 psi |